A. Introduction:
The YM-E0301 Discoloration Meter possesses oven, aging and anti-yellowing functions. A 300W UV lamp install inside the oven to simulate the fading or discoloring of sample caused by radiation of sunlight or inside high temperature containers during transportation.
B. Photo reference:

C. Specification:
1. Temperature range:Normal +5℃ ~ 260℃ (Generally 50℃)
2. Control method:Electric Digital temperature controller (With dual thermo protector)
3. Temperature display:0.1℃。
4. Heating method:Hot air recycling
5. Timer:0~999Hours, manual setting, digital display, automatically stop heating
6. Rotary speed of turning table:3 r.p.m. (Adjustable)
7. UV irradiance level:Type of Bulb OSRAM 300W. With counter for bulb's life. Life cycle for each bulb is about 1,000hours.
8. Accessories:Two pieces of shelf. Height adjustable
9. Internal material:Stainless Steel SUS #304
Outer material :Professional Pre-painting
10. Dimension:93×58×117cm (W×D×H)
11. Voltage:Single phase 220V/50Hz or 60Hz
12. Weight:100 kg。